> DDS Function Signal Generator
> Digital Multimeters
> Distortion Analyzer
> Four Probe experiment
> Function Generator
> IC Testers
> LCR Meters
> Oscilloscopes
> Power Supplies
> Pulse Wave Frequency Meter
> Signal / Function Generator
> Storage Oscilloscopes
> Two Probe Setup
> Withstand Tester
> General Physics Equipments
> TEKON INSTRUMENTS
  - T100 Network Cable Detector
  - T550 Power Quality Analyzer
  - T560 Power Quality Analyzer
  - T600 Txr Turn to Ration
  - T610 Txr Analyzer
  - T650 ESS Diagnostic
  - T700 Ind. Robot Diagnostic
  - T800 EV-HEV Diagonstic
 
   
   
   
Withstand Tester Hipot Tester

Withstand Tester

Hipot Tester

Ground Bond Tester

2500 Series Hipot Testers

Relevant Applications
Motors • Switches and Controls • Manufactured / Modular Homes • Recreational Vehicles
 
Features
Safety Made Simple
• Portable design and user-friendly interface
• Remote Safety Interlock
• Optional Ground Continuity (2503, 2510, 2550)
• Tamper proof front panel controls
   
       
2503 - DC Hipot  
The 2503 is a 5 KV, 3 mA DC Hipot tester that is designed for quick and easy production line DC Hipot testing. It features a LED display, menu driven parameter control, electronic ramp and dwell and a memory or parameter storage. An optional ground continuity check feature is available.
   
2510 - AC Hipot  
The 2510 is a 5 KV, 10 mA AC Hipot tester that is designed for quick and easy production line testing. Features include an LED display, menu driven parameter control, electronic ramp and dwell and a memory for parameter storage. An optional ground continuity check feature is available.
   
2525 - AC Hipot, Ground Continuity, 500VA  
   
   
2550 - AC Hipot, DC Hipot  
The 2550 is a 5 KV, 10 mA AC and a 5 KV, 3 mA DC Hipot tester that is designed for quick and easy production line Hipot testing. It features a LED display, menu driven parameter control, electronic ramp and dwell and a memory for parameter storage. An optional ground continuity check feature is available.
 
Model 2503 2510 2525 2550
  DC Hipot AC Hipot AC Hipot, Ground Continuity, 500VA AC Hipot, DC Hipot
Input
Voltage

115/230V selectable +/- 15%

115/230V selectable +/- 15%

-

 

115/230V selectable +/- 15%
Frequency 47 ~ 63 Hz 47 ~ 63 Hz - 47 ~ 63 Hz
Fuse 115VAC, 230VAC 2A fast acting 250VAC 115VAC, 230VAC 2A fast acting 250VAC 115VAC, 230VAC 15A fast acting 250VAC 115VAC, 230VAC 2A fast acting 250VAC
         
 
Dielectric Withstand Test mode
Output DC 0 – 5000 V, 3 mA AC 0 – 5000 V, 10 mA AC 0 – 5000V, 10V / step, 100mA AC 0 – 5000 V, 10 mA; DC 0 – 5000 V, 3 mA
- Regulation ± (1 % of output + 5V)
Voltage Setting
- Range 0.01 kV – 5 kV 0.01 kV – 5 kV - 0.01 kV – 5 kV
- Resolution 10 volts/step
- Accuracy ± (2 % of Setting + 5 V) (relative to displayed output). Can be adjusted during operation via Up & Down Arrow keys.
Ripple < 5 % at 5 kVDC / 3 mA - - < 5 % at 5 kVDC / 3 mA
Dwell Time 0, 1 or 60, 0 for continuous running 0, 1 or 60, 0 for continuous running 0 and 0.2 – 999.9 seconds, 0.1 second /
step “0” for continuous running
0, 1 or 60, 0 for continuous running
Ramp Timer 0 and 0.2 - 999.9 sec: 0.1 sec / step 0 and 0.2 - 999.9 sec: 0.1 sec / step - 0 ramp setting: 0.1 sec fixed ramp
Failure Setting
- High Limit 0.02 – 3.00 mA 0.10 – 10.00 mA 0.00 – 99.99 mA 0.10 – 10.00 mA and 0.02 – 3.00 mA
- Low Limit - - 0.00 – 99.99 mA, 0.01 mA/
step (0= OFF)
-
- Resolution 0.01 mA / Step
- Accuracy ± (2 % of setting + 0.02 mA)
Metering        
- Voltmeter Range DC 0.00 – 5.00 kV AC 0.00 – 5.00 kV AC 0.00 – 5.00 KV AC 0.00 – 5.00 kV; DC 0.00 – 5.00 kV
- Resolution 0.01 kV
- Accuracy ± (2 % of reading + 10 V)
- Ammeter Range DC 0.00 – 3.00 mA AC 0.00 – 10.00 mA AC 0.00 - 99.99 mA AC 0.00 – 10.00 mA; DC 0.00 – 3.00 mA
- Resolution 0.01 mA 0.01 mA 0.01 mA 0.01 mA
- Accuracy ± (2 % of reading + 0.02 mA)
Timer Display
- Range 0.0 – 999.9 seconds
- Resolution 0.1 second
- Accuracy ± (0.1% of reading + 0.05 sec)
Discharge Time ≤ 300 ms     ≤ 300 ms
Max. Capacitive Load 1 uF < 1 kV
0.75 uF < 2 kV
0.5 uF < 3 kV
0.08 uF < 4 kV
0.04 uF < 5 kV
    1 uF < 1 kV
0.75 uF < 2 kV
0.5 uF < 3 kV
0.08 uF < 4 kV
0.04 uF < 5 kV
Optional Ground Continuity Check
- Current DC 0.1 A ± 0.01 A, fixed
- Max Ground Resistance 1Ω ± 0.1 Ω, fixed
 
General Specifications
PLC Remote Control and Signal Output

The following input & output signals are provided through the 9 pin D type connector: 1. Remote control: test, reset & interlock. 2. Outputs: pass, fail & test in process.

Line Cord Detachable 6’ (1.80 m) power cable terminated in a three prong grounding plug.
Terminations 6’ (1.80 m) high voltage safety retracting probe, 6’ (1.80 m) high voltage clip probe & 6’ (1.80 m) return clip lead.
Dimensions (W x H x D) 4.75” x 5.75” x 14.50”
(120 x 146 x 370 mm)
4.75” x 5.75” x 14.50”
(120 x 146 x 370 mm)
17” x 5.8” x 12”
(432 x 147 x 305mm)
4.75” x 5.75” x 14.50”
(120 x 146 x 370 mm)
Weight 16.5 lbs. (7.5 kgs) 16.5 lbs. (7.5 kgs) 56.70 Lbs. (25.92 kgs) 16.5 lbs. (7.5 kgs)
         
 

2900 Series Hipot Testers

Relevant Applications
• Appliance • Lighting & LED Testing • Power Tools • Motors & Pumps
 
Features:
• Intuitive user interface
• 10 programmable memories
• Tamper proof front panel controls
• Includes Continuity test function
• Automate with PLC control or optional RS-232 interface
• Remote Safety Interlock
   
2955 - AC Hipot, Ground Continuity
Featuring our most advanced technology, the 2955 is a 5 KV, 12 mA AC Hipot tester that allows manufacturers to quickly and cost-effectively perform AC Hipot tests. It includes features such as a liquid crystal display, menu driven parameter control, electronic ramp, fully adjustable electronic dwell, fully adjustable Continuity test mode, 10 test set up memories, front panel lockout, standard remote safety interlock and an optional RS-232 interface. The 2955 is also NRTL safety agency listed.
   
2965 - AC Hipot, DC Hipot, Ground Continuity
Featuring our most advanced technology, the 2965 is a 5 KV, 12 mA AC and 6 KV, 5 mA DC Hipot tester that allows manufacturers to quickly and easily perform AC or DC Hipot tests. It includes features such as a liquid crystal display, menu driven parameter control, electronic ramp, fully adjustable electronic dwell, fully adjustable Continuity test mode, 10 test set up memories, front panel lockout, standard remote safety interlock and an optional RS-232 interface. The 2965 is also NRTL safety agency listed.
   
2975 - AC Hipot, DC Hipot, Ground Continuity, Insulation Resistance
Featuring our most advanced technology, the 2975 is a 3-in-1 Hipot tester that will perform AC, DC Hipot with a built-in Insulation Resistance tester. It will perform a 5 KV, 12 mA AC and a 6 KV, 5 mA DC Hipot test. It includes features such as a liquid crystal display, menu driven parameter control, electronic ramp, fully adjustable electronic dwell, fully adjustable electronic dwell, fully adjustable Continuity test mode, 10 test set up memories, front panel lockout, standard remote safety interlock and an optional RS-232 interface. The 2975 is also NRTL safety agency listed.
   
Model 2955 2965 2975
  AC Hipot, Ground Continuity AC Hipot, DC Hipot, Ground Continuity AC Hipot, DC Hipot, Ground Continuity, Insulation Resistance
Input      
Voltage

115/230V selectable +/- 15%

Frequency 50/60 Hz ± 5%
Fuse 2 A / 250 V fast acting 250 VAC
Dielectric Withstand Test mode
Output
- AC AC 0 - 5000 V, 2 V/step, 12 mA    
- DC - DC 0 - 6000 V, 2 V/step, 5 mA DC 0 - 6000 V, 2 V/step, 5 mA
- Regulation ± (1% of output + 5 V)
Voltage Setting
- Range 0 V - max output rating, 10 V/step
- Resolution 0.01 kV
- Accuracy ± (2% of setting + 5 V) relative to displayed output, can be adjusted during operation via UP and DOWN arrow keys
AC Output Frequency 50/60 Hz user selectable
AC Wave Form Sine Wave
AC Distortion <2% THD
DC Ripple - <5% at 6 kVDC / 5 mA <5% at 6 kVDC / 5 mA
Dwell Time - Range 0 and 0.2 - 999.9 seconds
0.1 sec/step
0 for continuous running
Ramp Timer - Range 0 and 0.2 - 999.9 seconds
0.1 sec/step
0 ramp setting = 0.1 sec fixed ramp
2955: ramp = 0.1 sec fixed ramp
AC Mode Failure Setting
- High Limit 0.10 - 12.00 mA, 0.01 mA/step
- Low Limit 0.00, 0.10 - 12.00 mA, 0.01 mA/ step (0=OFF) ( 2955, 2965, and 2975)
- Accuracy ± (2% of setting + 0.02 mA)
DC Mode Failure Setting
- High Limit 0.02 - 5.00 mA, 0.01 mA/step
- Low Limit 0.00, 0.02 - 5.00 mA, 0.01 mA/step
(0=OFF) ( 2965 and 2975)
- Accuracy ± (2% of setting + 0.02 mA)
Timer Display
- Range 0.0 – 999.9 sec
- Resolution 0.1 sec
- Accuracy ± (0.1% of reading + 0.05 sec)
Discharge Time ≤ 200 ms
Max. Capacitive Load in DC Mode 1.00 uF <1kV 0.08 uF < 4kV
0.75 uF <2kV 0.04 uF < 5kV
0.50 uF <3kV 0.01 uF < 6kV
Optional Ground Continuity Check
- Current DC 0.1 A ± 0.01 A, fixed
- High Limit 0.00-1.50 Ω
- Low Limit 0.01-1.50 Ω
- Offset 0.00-0.50 Ω
- Resolution 0.01 Ω
- Accuracy ± (2% of setting + 0.02 Ω)
 
Insulation Resistance Test Mode
Output Voltage  
- Range 100 – 1000 VDC
- Resolution 10 V/step
- Accuracy ± (2% of reading + 5 volts)
Voltage Display  
- Range 100 – 1000 V
- Resolution 10 V/step
- Accuracy ± (2% of reading + 5 volts)
Resistance Display  
- Range 1–1000 MΩ (4 digit, auto ranging)
- Resolution   500VDC 1000VDC
 
  0.01 1.00–40.00 1.00–80.00
  0.1 35.0–999.9 75.0-999.9
- Accuracy ± (3% of reading + 2 counts) at test voltage > 500 V
± (7% of reading + 2 counts) at test voltage ≤ 500 V
High Limit Range 0 - 1000 MΩ (0 = off)
Low Limit Range 1 - 1000 MΩ
Delay Timer  
- Range 0, 0.5 – 999.9 sec (0 = constant)
- Resolution 0.1 sec
- Accuracy ± (0.1% of reading + 0.05 sec)
   
 
General Specifications
Remote Control and Signal Output The following input and output signals are provided through the 9-pin D-type connector:
1. Remote control: Test, Reset & Interlock
2. Outputs: Pass, Fail, Test-in-Process
Memories 10 memories
Security Lockout capability to avoid unauthorized access to all test parameters. Memory lockout to avoid unauthorized access to memory locations.
Interlock Remote safety interlock standard.
Calibration Software and adjustments made through front panel.
Line Cord Detachable 6’ (1.80 m) power cable terminated
in a three prong grounding plug
Interface Optional RS-232 interface
Mechanical Tilt up front feet
Dimensions (W x H x D) 11” x 3.5” x 11.81” (280 x 89 x 300 mm)
Weight 20 lbs. (9 kg)
       
 

1300 Series Hipot Testers

Relevant Applications
• Motors • Manufactured / Modular Homes • Recreational Vehicles
 
Features
• Portable design and user-friendly interface
• Exclusive PULSE mode for continuous testing
• Remote Safety Interlock (1340)
• Optional Ground Continuity (1340)
 
 
 
Model 1305 1340
  AC Hipot, 3kV AC Hipot, 2.5 kV
Input    
Voltage

115/230V AC +/- 15%, user selection

Frequency 47 - 63 Hz
Fuse 115VAC, 230 VAC 3.15A slow blow 250 VAC
Test mode
Output
-AC 0 - 3000 V, 5 mA 2.5 kVAC @40mA;
    2.5 kVAC @10mA when pulse mode is activated
-Regulation ± (1% of output + 5 V) from no load to full load
Voltage Setting
- Range 0.01 kV - 3.00 kV 0.00 - 2.50 kV
- Resolution 10V/step
- Accuracy ± (2% of setting + 5 V) relative to displayed output
Dwell Setting On, HV remains on after “TEST” button is pushed 0.0 sec, 1.0 sec, 60.0 sec, off
    0.0 for continous running
  Off, HV present only when “TEST” button
is pushed
OFF: HV present only when “TEST” button
is pushed
Ramp/ Pulse Setting On, Automatic reset after failure for
continuous testing. (Not for Compliance
Testing)
PULSE: automatic reset after failure for
continous testing.
    Range: Pulse, 0.1 – 999.9 sec
    Resolution: 0.1 second/step
    Accuracy: ± (0.1% + 0.05 sec)
  Not for compliance testing
Failure Setting    
-High Limit 0.00 – 5.00 mA 0.00 – 40.00 mA,
    0.00 – 10.00 mA when PULSE mode is activated
-Resoultion 0.01 mA / Step 0.01 mA / step
-Accuracy ± (2 % of setting + 0.02 mA) ± (2% of setting + 0.02 mA)
 
Metering (4 digits)
Voltmeter  
- Range AC 0.00 – 3.00 kV 0.00 – 2.50 kV
- Resolution 0.01 kV 0.01 kV
- Accuracy ± (2 % of reading + 10 V) ± (2% of reading + 10 volts)
Ammeter  
- Range   0.00 – 40.00 mA,
    0.00 – 10.00 mA when PULSE mode is activated
-Resolution   0.01 mA
-Accuracy   ± (2% of reading + 0.02 mA)
Timer  
- Range   0.0 – 999.9 sec
- Resolution   0.1 sec
- Accuracy   ± (0.1% of reading + 0.05 sec)
General Specifications
Remote Control The following input and output signals are provided through the 9-pin D-type connector:
  Test Input Test, reset, and interlock
  Outputs: pass, fail, and test-in-process
Optional Ground Continuity Check Max Ground Resistance: 1 Ω ± 0.1 Ω, fixed
Calibration Software and adjustments made through front panel.
Mechanical Bench top model
Line Cord Detachable 6’ (1.80 m) power cable terminated
in a three prong grounding plug
Terminations 6’ (1.80 m) high voltage safety retracting probe
  6’ (1.80 m) high voltage clip probe
  6’ (1.80 m) return clip lead
  Tilt up front feet
Dimensions (W x H x D) 4.75” x 5.75” x 14.5” (120 x 146 x 370 mm)
Weight 16.0 lbs. (7.25 kgs) 20 lbs. (9 kg)
     
 

2200 Series Hipot Testers

The 2205 is stand-alone Insulation Resistance tester designed for use on the production line or in the field. With measurements up to 200 GΩ at voltages up to 1000 VDC, the 2205 can satisfy even the most demanding application requirements. We’ve incorporated the most user-friendly interface in the industry and a portable design for safe and easy testing.
 
Relevant Applications
• Motors • Switches and Controls • Cable and Harness • Aerospace
 
Features
• Portable design
• User-friendly interface
• Automate with PLC control
• Remote Safety Interlock
 
 
Model 2205
  Insulation Resistance
Input  
Voltage

115/230V selectable +/- 15%

Frequency 50/60 Hz ± 5%
Fuse 1A / 250 V fast acting
Insulation Resistance Test
Output Voltage
- Range 30 - 1000 VDC
- Resolution 1 Volt
- Accuracy ± (1% of Setting + 1 V) (relative to displayed output)
- Ripple < 2%
Voltage Display
- Low Range 0.0 V - 100.0 VDC
- High Range 101 V - 1000 VDC
- Resolution 0.1 V (low range), 1 V (high range)
- Accuracy ± (2% of reading + 2 V)
Resistance Display  
-Range 0.01 MΩ - 200.0 GΩ (4 digit, auto ranging)
-Resolution   300-499V 500-1000V
  .001 MΩ - .1 MΩ 0.1 MΩ – 1 GΩ 0.1 MΩ – 1 GΩ
  1 MΩ - .01 GΩ 1 GΩ – 20 GΩ 1 GΩ – 20 GΩ
  0.1 GΩ   20 GΩ – 200 GΩ
-Accuracy  
    300-499V 500-1000V
    0.1MΩ-1GΩ ± (3% of reading + 2 counts) 0.1MΩ-1GΩ ± (2% of reading + 2 counts)
    1-20GΩ ± (5% of reading + 2 counts) 1-20GΩ ± (3% of reading + 2 counts)
      20-200GΩ ± (10% of reading + 2 counts)
Timer Display  
-Range 0.0 - 999.9 seconds
-Resolution 0.1 second
-Accuracy ± (0.1% of reading + 0.05 seconds)
Failure Settings  
-Low Limit 0.1 MΩ - 999.9 MΩ
  1000 MΩ - 9999 MΩ
  10.0 GΩ - 200.0 GΩ
Dwell Time Setting 1.0 - 999.9 seconds, 0.1 second / step
  “0” for continuous running
Delay Time Setting 0.1 - 999.9 seconds, 0.1 second / step
Discharge Automatic Discharge of Device Under Test
  Indicator: Green < 30 V, Red > 30 V
General Specifications
PLC Remote Control The following input and output signals are provided through the 9-pin D-type connector:
  1. Inputs: test , reset, safety interlock
 

2. Outputs: pass, fail and test in progress

Line Cord Detachable 6’ (1.80 m) power cable terminated in a three prong grounding plug
Terminations High Voltage Output: Alden Socket
  Shielded Return: BNC Connector
Dimensions (W x H x D) 4.75” x 5.25” x 11.75” (120 x 133 x 300 mm)
Calibration Software & adjustments made through front panel
Weight 11 lbs. (5 kg)
 
* Products not listed can be ordered as per request.